medical Class II Updated 2020-05-20

Covidien Llc recalls Dover 100% Silicone Premium Urine Meter Foley Tray, 400mL, 1

Recalled Product

Dover 100% Silicone Premium Urine Meter Foley Tray, 400mL, 14 Fr/Ch (4.7 mm), 5 mL Catheter Pre-connected to 2000 mL Drainage Bag with Luer-Lock Sampling, Temperature Sensor and Adhesive Catheter Securement, Sterile, Product Code: P4P14TSD - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.


Hazard / Issue

Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot Numbers: 1909308464, 1933000164
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →