medical Class II Updated 2023-06-28

Olympus Corporation of the Americas recalls Single Use Repositionable Clip QUICKCLIP PRO 230 CM 10/BX- I

Recalled Product

Single Use Repositionable Clip QUICKCLIP PRO 230 CM 10/BX- Intended for hemostatic metal clip for the GI Tract Catalog Number: HX-202UR.A


Hazard / Issue

Complaints on clip deployment occurring during clinical procedures, leading to hazardous situations where clips fail to achieve their expected function or deploy incorrectly, potentially causing harms to patients. The complaints include reports that: 1) the clip arm does not open when the user pushes the slider, 2) the clip arm does not close when the user pulls the slider and 3) the clip detaches from the target tissue earlier than expected, after being deployed in a procedure.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI: 04953170385933 Lot Numbers: 24K, 25K, 26K, 27K, 28K, 29K, 2XK, 2YK, 2ZK