medical MODERATE Updated 2026-05-06

Siemens Healthcare Diagnostics Inc recalls epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis

See all recalls from Siemens Healthcare Diagnostics Inc →

Recalled Product

epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

Use of the affected product can result in discrepant low pH and measured Total Carbon Dioxide (mTCO2) results and corresponding calculated values in arterial, venous, and capillary patient samples. An erroneously low pH or mTCO2 reading can lead to inappropriate treatment of nonexistent acidosis (respiratory of metabolic) or missed diagnosis of alkalosis (a true high pH: respiratory or metabolic alkalosis), potentially causing harm through unnecessary interventions with the former or failure to deliver appropriate treatment with the latter.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Siemens Material Number (SMN): 10736515; UDI-DI: 00809708121860; Lot numbers: 01-26024-70, 01-26026-70, 01-26028-70, 01-26030-70 (OUS only), 02-26027-80, 02-26028-80 (OUS only);