medical Class II Updated 2020-05-20

Ra Medical Systems, Inc. recalls Ra Medical Systems DABRA Catheter 5F (1.5mm), REF: 1222-5000

Recalled Product

Ra Medical Systems DABRA Catheter 5F (1.5mm), REF: 1222-5000-01, LOT: 18-090400-094, USE BY: 2019-09-24, MFR DATE: 2018-09-24, CE, Laser Radiation, STERILE (Gamma Sterilized), Single Use Only Revised Label: Product Label - Ra Medical Systems DABRA Catheter 5F (1.5mm), REF: 1222-5000-01, LOT: 19-092400-041, USE BY: 2019-12-13, MFR DATE: 2019-10-14, CE 2797, Laser Radiation, STERILE (Gamma Sterilized), Single Use Only Package Label - DABRA 101 5 French, Ref: 1222-5000-01, Lot: LOT: 19-092400-041, USE BY: 2019-12-13, UDI's (01)00842362100031, (17)20191213(21)2019092400041


Hazard / Issue

The firm has become aware there is a potential problem with its DABRA catheters resulting in higher catheter calibration failure rate due to the product shelf life is currently less than the labeled shelf life.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: DABRA 101 Catheter. Ref: 1222-5000-01 Serial Number(s): Manufacturing dates 2018-09-03 to 2019-08-26. Expiration dates: September 2019 to August 2020