Siemens Medical Solutions USA, Inc recalls AXIOM Artis MP, Model Number 5904466

Recalled Product

AXIOM Artis MP, Model Number 5904466


Hazard / Issue

A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Serial Numbers: 20334 20052 20303 20238 20242 20322 20282 20363 20338 20295 20348 20172 20264 20367 20258 20265 20250 20249 20281 20243 20339 20321 20294 20254