medical Class II Updated 2019-07-31

Siemens Medical Solutions USA, Inc recalls Artis zee floor MN, Model Number 10094142

Recalled Product

Artis zee floor MN, Model Number 10094142


Hazard / Issue

A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Serial Numbers: 140212 140115 140101 140205 140213 140204 140209 140219 140224 140218 140221 140220 140206 140223 140105 140226 140108 140102 140109 140100 140202 140214 140114