medical Class II Updated 2020-05-27

Cardinal Health Inc. recalls B-Hydroxybutyrate LiquiColor, Item code CH2440058 - Product

Recalled Product

B-Hydroxybutyrate LiquiColor, Item code CH2440058 - Product Usage: Product is used for the quantitative determination of B-Hydroxybutyrate in human serum or plasma.


Hazard / Issue

Three products requiring refrigerated storage conditions were incorrectly stored outside of approved temperature ranges. Using these temperature abused products may result in misdiagnosis of the patient.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot 193462
View official government recall