medical Class II Updated 2019-07-31

Zyno Medical LLC recalls Zyno Medical Administration Set -105"ADMIN SET 20DP, W/2 Y S

Recalled Product

Zyno Medical Administration Set -105"ADMIN SET 20DP, W/2 Y SITE, CVCLMPS- For Use with Zyno Medical Infusion pumps or Gravity Feed Product Code:B2-70072


Hazard / Issue

Administration sets assembled with inverted pinch clamp cause the fluid from the distal end of the administration set to be pumped to the proximal end of the tubing when secondary set is used with the primary set. Thus, instead of delivering fluid to the patient, blood may be drawn from the patient if there is space in the drug container to hold extra volume of fluid, under the condition that the reversed pinch clamp is not identified

Issued by

FDA

Affected States: MA
Lot/Code Info: Lot Numbers:18037128,18065375, 19015969 UDl/UPC/GTIN: 00814371020198