medical Class II Updated 2019-07-31

LABORIE MEDICAL TECHNOLOGIES, CORP recalls U3-3 Solar Blue, Laptop integrated power supply used with th

Recalled Product

U3-3 Solar Blue, Laptop integrated power supply used with the Solar Blue Urodynamics for assessing lower urinary tract dysfunctions


Hazard / Issue

Power supply unit (PS-SBL) not correctly labeled, and insulation material used at the low voltage DC end does not meet all applicable flammability rating requirements. may lead to a potential fire hazard

Issued by

FDA

Affected States: AZ, PA
Lot/Code Info: Serial Numbers: PE18-7SBL1452, PE18-7SBL 1414