medical Class III Updated 2019-07-31

Abbott Laboratories recalls ARCHITECT BNP Controls Ptoduct Usage: The ARCHITECT BNP

Recalled Product

ARCHITECT BNP Controls Ptoduct Usage: The ARCHITECT BNP Controls are for the estimation of test precision and the detection of systematic analytical deviations of the ARCHITECT iSystem (reagents, calibrators and instrument), when used for the quantitative determination of human B type natriuretic peptide (BNP) in human EDTA plasma.


Hazard / Issue

Presence of a time dependent, stability drift in patient and control results returned from ARCHITECT BNP testing.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: a. List Number 8K28 11 - Lot Number (Expiration Date) UDI: 44K78918 (18JUN2019) (01)00380740017477(17)190618(10)44K78918, 44K80918 (05SEP2019) (01)00380740017477(17)190905(10)44K80918, 44K82718 (15NOV2019) (01)00380740017477(17)191115(10)44K82718; b. List Number 8K28 12 - Lot Number (Expiration Date) UDI: 44K79518 (23Jul2019) (01)00380740145262(17)190723(10)44K79518, 44K81118 (5Sep2019) (01)00380740145262(17)190905(10)44K81118, 44K82818 (15Nov2019) (01)00380740145262(17)191115(10)44K82818, 44K84219 (17Dec2019) (01)00380740145262(17)191217(10)44K84219
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →