medical Class II Updated 2019-08-07

Encore Medical, Lp recalls Flexible Drill Shaft, FMP Acetabular System Instrumentation

Recalled Product

Flexible Drill Shaft, FMP Acetabular System Instrumentation REF 803-05-239 RX Encore Medical, L.P.


Hazard / Issue

Manufacturing error, the drill bit shaft is unable to properly capture the drill bit pin. Failure of the drill bit shaft to capture the drill bit pin could result in the drill bit coming loose which could result in a delay in surgery or adverse patient reaction resulting in a revision surgery.

Issued by

FDA

Affected States: MS, NY, MO, OK, UT, AZ
Lot/Code Info: Lot # 262667L01
View official government recall