medical Class II Updated 2021-07-28

Gentherm Medical, LLC recalls Hemotherm CE Dual Reservoir Cooler/Heater, Model 400CE, 115V

Recalled Product

Hemotherm CE Dual Reservoir Cooler/Heater, Model 400CE, 115V Model, part no. 86022. Used to cool or warm blood during cardiopulmonary bypass procedures


Hazard / Issue

There is a potential risk of device contamination and patient infection associated with the device because of the potential for organisms (including Nontuberculous mycobacteria NTM) to grow in the water systems of any heater-cooler device, and contaminated water from any heater-cooler device has the potential to aerosolize into the operating room during surgery which could lead to patient infection.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: All units in service life Serial Numbers: 092-10011CE tot 212-11499CE
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →