medical Class I Updated 2021-08-18

Baxter Healthcare Corporation recalls Spectrum IQ Infusion System with Dose IQ Safety Software, Pr

Recalled Product

Spectrum IQ Infusion System with Dose IQ Safety Software, Product Code 3570009. for the controlled administration of fluids.


Hazard / Issue

System errors were noted following changes the customers implemented to the configuration of their network and server systems. Gateway Server System performance anomalies resulted in excessive Spectrum IQ pump connectivity errors.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: All Serial Numbers
View official government recall

Looking for safe alternatives?