medical Class II Updated 2018-06-13

Orthofix, Inc recalls CC Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label,

Recalled Product

CC Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) Small (002-G-01020); b) Medium (002-G-01025); c) Large (002-G-01030)


Hazard / Issue

Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.

Issued by

FDA

Affected States: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK
Lot/Code Info: (UDI)/Lot: a) (18056099647694) B1099067; b) (18056099647700) B1099061; c) (18056099647717) B1099063