medical Class II Updated 2023-07-19

Ortho-Clinical Diagnostics recalls VITROS Chemistry Products Calibrator Kit 20-in vitro diagnos

Recalled Product

VITROS Chemistry Products Calibrator Kit 20-in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 20- Used to calibrate the VITROS 5,1 FS/4600 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems for the quantitative measurement of transferrin, C3, C4, IgA, IgG and IgM. Product Code: 6801704


Hazard / Issue

Potential to cause biased results in the upper end of the reportable range for the quantitative measurement of Transferrin, C3, C4, IgA, IgG and IgM. Within the reference interval, the highest bias observed in selected sample testing investigation was +35.3% for IgM; -4.3% for IgG; +12.3% for IgA; -6.9% for C3; -6.6% for C4; and -17.8% for transferrin. Falsely elevated or reduced results could be reported depending on the analyte and/or sample concentration.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI: 10758750006687 Lot Code: 2022 Exp. Date :06-Sept-2023