medical Class II Updated 2018-06-13

Orthofix, Inc recalls L-Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, L

Recalled Product

L-Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Left or Right (Part Number): a) Left (003-E-00001L); b) Right (003-E-00001R)


Hazard / Issue

Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.

Issued by

FDA

Affected States: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK
Lot/Code Info: (UDI)/Lot: a) (18056099647830) B1108952; b) (18056099647847) B1108955
View official government recall