medical Class II Updated 2020-06-03

Beckman Coulter Inc. recalls ISE Electrolyte Reference (2 X 2L) Reagent, REF A28937, UDI:

Recalled Product

ISE Electrolyte Reference (2 X 2L) Reagent, REF A28937, UDI: 15099590202491 - Product Usage: are intended for the quantitative determination of carbon dioxide in human serum or plasma.


Hazard / Issue

Electrolyte reference reagent, used for clinical laboratory testing to determine CO2 in human serum or plasma for diagnosis and treatment of potentially serious disorders associated with changes in the body acid-base balance, has a low pH and may cause CO2 calibration failures, high CO2 quality control recoveries, and false high patient results which can lead to delay in treatment.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot Number: M909206
View official government recall