medical Class II Updated 2021-08-04

B. Braun Medical, Inc. recalls Infusomat Space Volumetric Pump Administration Set, Product

Recalled Product

Infusomat Space Volumetric Pump Administration Set, Product Code 490100


Hazard / Issue

There is a potential for the Anti-free flow clip of the administration set to be inverted which may cause unintended free flow if the user fails to follow on screen prompt to close the roller clamp prior to set removal from the pump.

Issued by

FDA

Affected States: CA, MI, OR, UT
Lot/Code Info: Lot Number: 0061761278
View official government recall