medical Class II Updated 2020-06-03

Radiometer Medical ApS recalls ABL90 FLEX PLUS, model no. 393-02; and Spare Part Hemolyzer

Recalled Product

ABL90 FLEX PLUS, model no. 393-02; and Spare Part Hemolyzer Units (902-980): R0526 - R0941 - Product Usage: are intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting.


Hazard / Issue

The analyzers fail on S7950 (Range +) QC measurements due to the hemolyzer cuvettes, a subcomponent of the optical system which measures ctHb and derivates. In a cuvette with light shunt, some of the received light has not been transmitted through the sample. This will lead to a perturbed and non-linear absorbance spectrum which results in measurement errors.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: ABL90 FLEX PLUS Model no. 393-092: R0014 - R0195; and Spare Part Hemolyzer Units (902-980): R0526 - R0941; Model Number 393-092.
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →