medical Class III Updated 2018-06-20

Alere San Diego Inc. DBA Immunalysis Corporation recalls lmmunalysis EDDP Urine Control Set 1 The EDDP Urine Contr

Recalled Product

lmmunalysis EDDP Urine Control Set 1 The EDDP Urine Controls (High 125 ng/ml and Low 75 ng/ml) are used as control materials in the ammonolysis EDDP Specific Urine Enzyme Immunoassay, which has a cutoff of 100 ng/ml. This test provides only a preliminary analytical test result and a more specific alternate chemical method, such as Gas Chromatography/Mass Spectrometry or Liquid Chromatography/Tandem Mass Spectrometry, must be used to obtain a confirmed analytical result.


Hazard / Issue

The High (125 ng/ml) Control does not meet specification - it has been confirmed to produce values below 100 ng/mL when run on validated clinical chemistry analyzers, which is below the 100 ng/mL cutoff of the EDDP Specific Urine Enzyme Immunoassay with which this control is utilized.

Issued by

FDA

Affected States: CA, CO, CT, FL, LA, MD, NC, NJ, OH, OK, TN, TX, WI, WV
Lot/Code Info: Part Number: 3006-5, Lot Number: EK16135