Hamilton Medical, Inc. recalls HAMILTON C2 Ventilator Is intended to provide positive pr
Recalled Product
HAMILTON C2 Ventilator Is intended to provide positive pressure ventilatory support to adults and pediatrics, and optionally infants and neonates
Hazard / Issue
Due to a software anomaly after approximately 91 days of cumulative use, the ventilator may stop and enter the ambient mode without prior notice.
Issued by
FDA
Distribution: U.S.: Alabama Alaska Arizona Arkansas California Colorado Connecticut Delaware Florida Georgia Hawaii Idaho Illinois Indiana Iowa Kansas Kentucky Louisiana Maine Maryland Massachusetts Michigan Minnesota Mississippi Missouri Montana Nebraska Nevada New Hampshire New Jersey New York North Carolina North Dakota Ohio Oklahoma Oregon Pennsylvania Rhode Island South Carolina South Dakota Tennessee Texas Utah Vermont Virginia Washington West Virginia Wisconsin Wyoming Added 08/30/2023 - Puerto Rico O.U.S.: Bahamas, Japan and Mexico