medical Class II Updated 2019-08-14

Howmedica Osteonics Corp. recalls Serrato 9.5mm X 90mm Polyaxial Screw Catalog Number 4826195

Recalled Product

Serrato 9.5mm X 90mm Polyaxial Screw Catalog Number 482619590 Serrato 9.5mm X 100mm Polyaxial Screw Catalog Number 4826195100


Hazard / Issue

Discrepant length; Lot B88751 90mm screws were manufactured with a 100mm length instead of the required 90mm length; lot B90243 100mm screws were manufactured with a 90mm length instead of the required 100mm length.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot # B88751 UDI: (01)07613327351576(11)190125(10)B88751