medical Class II Updated 2018-06-20

Smith & Nephew, Inc. recalls smith&nephew GENESIS(R) II TIBIAL Punch, 13 MM, REF 71440408

Recalled Product

smith&nephew GENESIS(R) II TIBIAL Punch, 13 MM, REF 71440408, QTY: (1)


Hazard / Issue

One lot of GENESIS II 13 mm tibial punches are 2 mm smaller than specification.

Issued by

FDA

Distribution: International only: United Arab Emirates, Spain, Mexico
Lot/Code Info: 15FM16181