medical Class II Updated 2021-08-04

Merit Medical Systems, Inc. recalls basixALPHA Inflation Device 30 atm/bar 20 mL - Product Usage

Recalled Product

basixALPHA Inflation Device 30 atm/bar 20 mL - Product Usage: Used to inflate and deflate an angioplasty balloon or other interventional device, and to measure the pressure within the balloon.


Hazard / Issue

Due to a design issue, inflation syringe handle separating from the inflation device when handle is withdrawn.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Catalog Number: IN3130 UDI: 00884450440179; 00884450510155; Lot Numbers: H1862928; H1896261; H1923761; H1929106; H1939255; H1966485; H1978295; H2068264; H2073193; H2073247; H2077173; Catalog Number: IN3152 UDI: 00884450511220 Lot Numbers: H1896262; H1909581; H1923762; H1929107; H1978299; H2007415; H2031599; H2068265; H2073248; H2077175; H2084368; Catalog Number: IN3302 UDI: 00884450511305 Lot Numbers: H1896263; H1966489; Catalog Number: IN3802 UDI: 00884450511510 Lot Numbers: H18962650; H1939259; H1966490; H1973390; H1978305; H1993118; H2001426; H2007417; H2012534; H2031601; H2073198; H2073249; H2077176;
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →