medical Class II Updated 2018-06-20

BioMerieux SA recalls The PREVI¿ Color Gram dyes are used with the PREVI¿ Color Gr

Recalled Product

The PREVI¿ Color Gram dyes are used with the PREVI¿ Color Gram instrument to stain bacterial and fungal microorganisms for direct examination, according to Gram s method. The PREVI¿ Color Gram instrument sprays the PREVI¿ Color Gram reagents onto microscope slides which have been smeared with specimens to be examined. The staining procedure is rapid and precise. Each reagent has a specific position (A, B, C, D, E) on the PREVI¿ Color Gram platform that is also indicated on the reagent label. An automated slide stainer is a device used to stain histology, cytology, and hematology slides for diagnosis.


Hazard / Issue

A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.

Issued by

FDA

Distribution: International distribution.
Lot/Code Info: Reagents for use with the PREVI¿ Color Gram: IODINE-B (ref: 29523) Lot number: G59407 Exp. 15-Aug-18 CRYSTAL VIOLET-C (ref: 29524) Lot number: G52306 Exp.19-Jul-18 FUCHSIN-A (ref: 29522): Lot Number Lot number: F89544 Exp. 13-Apr-18
View official government recall

Looking for safe alternatives?