medical Class II Updated 2021-08-11

BAXTER HEALTHCARE CORPORATION recalls Kaguya Automated Peritoneal Dialysis System

Recalled Product

Kaguya Automated Peritoneal Dialysis System


Hazard / Issue

Baxter Healthcare Corporation has identified that the AMIA Sharesource User Guide that is used with the AMIA Automated PD System cyclers incorrectly describes the function of the UF (Ultrafiltration) Limit setting as the ending criteria of the cyclers Extra Last Drain Mode. In some locations in the guide, it states that the UF volume achieved must exceed the programmed UF Limit before the cycler will end the Extra Last Drain, allowing the patient to end therapy or proceed to their last fill.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Product Code T5C8500 (distribution in Japan only)
View official government recall