medical Class II Updated 2018-06-20

Philips Medical Systems (Cleveland) Inc recalls Brilliance 16 with DEPMED HARDENING KIT 12NC: 453567400741,

Recalled Product

Brilliance 16 with DEPMED HARDENING KIT 12NC: 453567400741, Model Number 728240. Computed Tomography X-ray system


Hazard / Issue

The patient support head holder could contact the back ISO shelter wall during manual or motorized motion through the gantry bore. This may result in a serious injury that would require medical intervention to preclude permanent impairment in certain scenarios.

Issued by

FDA

Affected States: CA, MD, NV, OH, TX, WI
Lot/Code Info: System Serial Numbers: 6147, 6143