medical Class II Updated 2018-06-20

Cellavision AB recalls CellaVision DM Software versions 6.0.1 or 6.0.2 installed on

Recalled Product

CellaVision DM Software versions 6.0.1 or 6.0.2 installed on the following products: CellaVision DM96, DM1200, DM9600 and DI-60 Product Usage: The devices are automated cell-locating devices. The devices automatically locates and presents images of blood cells on peripheral blood smears. The operator identifies and verifies the suggested classification of each cell according to type. The devices are intended to be used by skilled operators, trained in the use of the device and in recognition of blood cells


Hazard / Issue

A software malfunction was found where WBC, RBC and PLT comments added after a slide is signed, are not sent to the LIS. This can only occur where customers process multiple slides per blood sample.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Versions 6.0.1 or 6.0.2
View official government recall