medical Class II Updated 2018-06-20

Roche Diabetes Care, Inc. recalls Accu-Chek Aviva Plus Test Strip, Blood glucose monitoring sy

Recalled Product

Accu-Chek Aviva Plus Test Strip, Blood glucose monitoring system, Catalog nos. 06908217001, 06908268001, 06908349001,06908217001 Product Usage: The Accu-Chek Aviva Plus test strips are for use with the Accu-Chek Aviva Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm.


Hazard / Issue

The recalling firm identified four US lots of test strips out of specification. Affected strips show increased potential for strip errors prior to applying blood, strips not detected once inserted, or biased results, which may not be detectable. In the case of a biased result, incorrect action could be taken by the user in response to the displayed blood glucose result.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lots 496947, 497291, 497296, 497325
View official government recall