medical Class II Updated 2021-08-11

Abbott Laboratories recalls REF 8L92-21, Abbott Activated Alanine Aminotransferase, for

Recalled Product

REF 8L92-21, Abbott Activated Alanine Aminotransferase, for use with Architect, IVD, CE, UDI: (01) 00380740017866


Hazard / Issue

There is a potential for >10% under-recovery on Activated Alanine Aminotransferase (A-ALT) samples greater than 1,200 U/L. In addition there is a potential for a >10% shift in patient results with samples containing bilirubin concentrations greater that 31 mg/dL. This could lead to incorrect patient results

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot Numbers / Expiration Date 15014UN20 20OCT2021 09153UN21 09MAR2022 *Not distributed within the US.
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →