medical MODERATE Updated 2024-07-03

Draeger Medical Systems, Inc. recalls Infinity Acute Care System (IACS) Monitoring System, Infinit

See all recalls from Draeger Medical Systems, Inc. →

Recalled Product

Infinity Acute Care System (IACS) Monitoring System, Infinity M500 with Power, Model No. MS20407. Rev 20, 21, and 23

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

Device is not in full compliance to Type CF requirements of IEC 60601-1 and IEC 60601-2-34 standards. Therefore, there is a potential risk of electric shock. Health consequences may include electrical shock, burns, and cardiac arrhythmia.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Model Number: MS20407. UDI-DI: 04049098054447. Rev 20, 21, and 23

✅ What Should You Do?

  1. 1Contact your healthcare provider immediately to discuss this recall.
  2. 2Stop using the device if it is safe to do so, and contact your provider.
  3. 3Follow the manufacturer's remedy instructions provided in the recall notice.
  4. 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.

Looking for a safe alternative?