medical Class II Updated 2020-06-10

Olympus Corporation of the Americas recalls OLYMPUS ShockPulse-SE Lithotripsy System, Model SPL-T Trans

Recalled Product

OLYMPUS ShockPulse-SE Lithotripsy System, Model SPL-T Transducer


Hazard / Issue

A component of the Olympus ShockPulse Transducer was manufactured with non-conforming material that could allow water ingress during autoclaving sterilization.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Serial numbers: 1910PW30607705, 1910PW30607712, 1910PW30607601, 1910PW30607708, 1910PW30607709, 1910PW30607701, 1910PW30607814, 1910PW30607906, 1910PW30607603, 1910PW30607706, 1910PW30607802, 1910PW30607808, 1910PW30607809, 1910PW30607804, 1910PW30607713, 1910PW30607803 Lot Numbers: CG2003, CG2010, CG2011, CG2026, CG2027, CG2033, CG2035, CG2000, CG2025, CG2001, CG2019, CG2022, CG2029, CG2021, CG2008, CG2036, CG2037, CG2006, CG2004, CG2007, CG2012, CG2013, CG2015