medical Class II Updated 2020-06-10

Inpeco S.A. recalls FlexLab Augomation System using the Aliquoter Module (Inpeco

Recalled Product

FlexLab Augomation System using the Aliquoter Module (Inpeco P/N FLX-212) The systems consolidate multiple Analytical instruments into a unified workstation.


Hazard / Issue

Potential for discrepant patient results due to a potential for the primary sample tube to be diluted. In certain firmware versions, in case a Clot Detection error (error code E0E0 or 13E0) is generated during the sample aspiration the current error recovery procedure dispenses 2/3 of sample volume back into the Primary Tube. Evidence from the field showed that in case of Clot Detection error, this management may lead to the dilution of the Primary Tube with the distilled water of the hydraulic circuit of the Aliquoter Module.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Serial Numbers: FLX.008 FLX.0105 FLX.0106 FLX.0107 FLX.0113 FLX.0114 FLX.0115 FLX.0116 FLX.0118 FLX.0119 FLX.012 FLX.0121 FLX.0124 FLX.0125 FLX.0126 FLX.0127 FLX.0128 FLX.0129 FLX.0131 FLX.0135 FLX.0136 FLX.0138 FLX.014 FLX.0140 FLX.0141 FLX.0143 FLX.015 FLX.0152 FLX.0155 FLX.0158 FLX.0161 FLX.0164 FLX.017 FLX.018 FLX.0196 FLX.020 FLX.021 FLX.022 FLX.023 FLX.027 FLX.029 FLX.030 FLX.031 FLX.032 FLX.033 FLX.037 FLX.038 FLX.039 FLX.043 FLX.044 FLX.046 FLX.047 FLX.050 FLX.051 FLX.052 FLX.056 FLX.060 FLX.064 FLX.071 FLX.072 FLX.074 FLX.075 FLX.076 FLX.077 FLX.080 FLX.082 FLX.083 FLX.093 FLX.101 FLX.102 TSH.003