medical Class II Updated 2019-08-14

Hycor Biomedical Inc recalls HYTEC 288 System, Model 74007P, automated immunoanalyzer (EI

Recalled Product

HYTEC 288 System, Model 74007P, automated immunoanalyzer (EIA) instrument. For in vitro diagnostic use only. HYTEC AUTOSTAT II RF assays: HYTEC AUTOSTAT II RF IgM Kit: Part Number FGA 05, HYTEC AUTOSTAT II RF IgA Kit: Part Number FGA 06, HYTEC AUTOSTAT II RF IgG Kit: Part Number FGA 07, HYTEC AUTOSTAT II Total Rheumatoid Factor Kit: Part Number FGA 08.


Hazard / Issue

Reports when using Rheumatoid Factor test kits, control and negative samples may be elevated following a sample containing high levels of IgG . The negative control may report out of range; a negative sample may recover as a low positive which could contribute to an incorrect or premature diagnosis of rheumatoid.

Issued by

FDA

Affected States: TX
Lot/Code Info: All lots of RF assay
View official government recall