Inpeco S.A. recalls Aptio Automation System using the Aliquoter Module (Inpeco P
Recalled Product
Aptio Automation System using the Aliquoter Module (Inpeco P/N FLX-212) - Product Usage: The systems consolidate multiple Analytical instruments into a unified workstation.
Hazard / Issue
In certain firmware versions, in case a Clot Detection error (error code E0E0 or 13E0) is generated during the sample aspiration the current error recovery procedure dispenses 2/3 of sample volume back into the Primary Tube. Evidence from the field showed that in case of Clot Detection error, this management may lead to the dilution of the Primary Tube with the distilled water of the hydraulic circuit of the Aliquoter Module.
Issued by
FDA