Agfa N.V. recalls AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dy

Recalled Product

AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic, Type number: 6010/200 - Product Usage: DR 800: Agfa s DR 800 system is indicated for performing dynamic imaging examinations (fluoroscopy and/or rapid sequence) of the following anatomies/procedures: positioning fluoroscopy procedures, gastro-intestinal examinations, urogenital tract examinations, and angiography. It is intended to replace fluoroscopic images obtained through image intensifier technology. In addition, the system is intended for project radiography of all body parts. The DR 800 is not intended for mammography applications


Hazard / Issue

Under specific conditions (Fluoroscopic exam, ABS=OFF, manual change of parameters) wrong calculation of the dose/minute for fluoroscopy exams can be possible

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Serial Numbers: 0925, 0945, 0948, 0949, 1003, 1022, 1116, 1136, 1161, 1229, 1239, 1401, and 1404