medical Class II Updated 2023-07-26

UFSK - International OSYS Gmbh recalls UFSK Treatment chair 500 XLE- Treatment chair for positionin

Recalled Product

UFSK Treatment chair 500 XLE- Treatment chair for positioning and support of the patient during diagnostic and therapeutic treatments of temporary to short-term duration, especially in ophthalmology Art.no. / REF no.: 15000004 Art.no. / REF no.: 15000006


Hazard / Issue

Service life of the electrical seat part drive of the treatment chair may be shortened unexpectedly after several years of use due to a potential wear-related reduction in the self-locking of the drive and may remain undetected. A worsened self-locking of the seat drive can lead to overriding the drive s end position when positioning the patient with the seat part in an inclined position

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Art.no. / REF no.: 15000004 Serial number range: 2000- 2999 Art.no. / REF no.: 15000006 Serial number range: 3000-4506
View official government recall