medical Class II Updated 2020-06-10

Zimmer Biomet, Inc. recalls Vital MIS Spinal Fixation system implants CANN POLY EXT TAB

Recalled Product

Vital MIS Spinal Fixation system implants CANN POLY EXT TAB 7.5X40MM, Item Number 810M7540 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle screw fixation, posterior hook fixation, or anterolateral fixation. Pedicle screw fixation is indicated for skeletally mature patients and for adolescence patients.


Hazard / Issue

There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI: (01)00887868325274; Lot Numbers: C19I0003 C19I0020 C19J0017 C19L0008 C20B0004 SBM108414 SBM118996 SBM120650 SBM122346