medical MODERATE Updated 2026-06-03

GE Medical Systems Information Technologies Inc recalls GE HealthCare APEXPRO v5.0 - ARK2250 - Software Disaster Rec

See all recalls from GE Medical Systems Information Technologies Inc →

Recalled Product

GE HealthCare APEXPRO v5.0 - ARK2250 - Software Disaster Recovery USB Kit - Orderable by GE Service only, REF #2108523-01; ApexPro CARESCAPE Telemetry Server (CTS) v5.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).

Issued by

FDA

Distribution: Worldwide distribution.
Lot/Code Info: UDI none, Serial number or Sales Order number SS921120014SA, SS921120017SA, SS921120021SA, SS921120018SA, SS921120020SA, SS920390001SA, SS921120007SA, SS921120008SA, SS921120012SA, SS921120016SA, SS920500001SA, SS920500002SA, SS921120009SA, SS921120010SA.

✅ What Should You Do?

  1. 1Contact your healthcare provider immediately to discuss this recall.
  2. 2Stop using the device if it is safe to do so, and contact your provider.
  3. 3Follow the manufacturer's remedy instructions provided in the recall notice.
  4. 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.

Looking for a safe alternative?

Search Amazon for GE HealthCare APEXPRO v5.0 →
View official government recall