GE Medical Systems Information Technologies Inc recalls GE HealthCare IOU ASSY ARK2250 CARESCAPE TELEMETRY SERVER, R
See all recalls from GE Medical Systems Information Technologies Inc →Recalled Product
GE HealthCare IOU ASSY ARK2250 CARESCAPE TELEMETRY SERVER, REF #2108729-901, Product Code SS9; ApexPro CARESCAPE Telemetry Server (CTS) v5.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Hazard / Issue
GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).
Issued by
FDA
✅ What Should You Do?
- 1Contact your healthcare provider immediately to discuss this recall.
- 2Stop using the device if it is safe to do so, and contact your provider.
- 3Follow the manufacturer's remedy instructions provided in the recall notice.