medical Class II Updated 2020-06-10

Abbott Laboratories recalls Accelerator APS Aliquoter Module, List number: 07L40 - Produ

Recalled Product

Accelerator APS Aliquoter Module, List number: 07L40 - Product Usage: The ACCELERATOR APS Workcell is a modular system designed to automate sample handling and processing in the clinical laboratory. The system allows consolidations of multiple clinical chemistry and immunoassay analytical instruments into a unified workstation.


Hazard / Issue

In the case of an AQM clot detection error a portion of the distilled water from the Aliquoter Module hydraulic circuit may be dispensed into the Primary Tube leading to sample dilution, which could lead to incorrect results.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: All serial numbers in the field
View official government recall

Looking for safe alternatives?