PENTAX of America Inc recalls C2 CryoBalloon Focal Catheter, REF FG-1009. C2 CryoBalloon
Recalled Product
C2 CryoBalloon Focal Catheter, REF FG-1009. C2 CryoBalloon Ablation System is used as a cryosurgical tool in the field of general surgery, specifically for endoscopic applications.
Hazard / Issue
The catheter alignment arrow printed onto the strain relief is not aligned with the detachment feature of the catheter connector, resulting in the potential for the catheter to be attached to the controller in the detach orientation. In this orientation, the catheter is not securely attached to the controller and may be partially (~1cm) ejected from controller.