TriMed Inc. recalls Nitinol Staple, Elastic Staple Small-Asymmetrical, Sterile R
Recalled Product
Nitinol Staple, Elastic Staple Small-Asymmetrical, Sterile R, Model #: a) ES-10x13x15; b) ES-10x16x19; c) ES-10x17x19; d) ES-12x13x15; e) ES-12x15x17; f) ES-12x16x19; g) ES-15x13x16
Hazard / Issue
During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not be found. Sterilization records indicate that all lots of staples were irradiated to at least the defined dose. There is a risk that this issue could result in patient infection if not properly sterilized.
Issued by
FDA