medical Class II Updated 2023-08-09

Maquet Cardiovascular, LLC recalls ACROBAT SUV Vacuum Off-Pump System, Part Number C-OM-9000S.

Recalled Product

ACROBAT SUV Vacuum Off-Pump System, Part Number C-OM-9000S. Component of a chest stabilization system for cardiac and chest surgery.


Hazard / Issue

The Standard Blade may not securely latch onto the Activator Drive. This may result in the inability to stabilize and position or may result in the loss of stabilization and positioning of heart, release of device component into the patient, and/or a procedural delay and/or conversion.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI: 00607567700543
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →