medical Class II Updated 2021-08-18

Abbott Laboratories, Inc recalls Alinity s System REF 06P16-01. An automated immunoassay an

Recalled Product

Alinity s System REF 06P16-01. An automated immunoassay analyzer for In Vitro diagnostic use in blood donor screening, plasma and plasmapheresis screening, and organ donor centers, hospitals, and reference laboratories.


Hazard / Issue

A design defect (hardware and software) allows liquid waste pressure to build up and potentially spray users/operators.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Model Number: LN 06P16-01 Serial numbers AS1001 through AS1330