medical Class II Updated 2020-06-17

Elekta Limited recalls Elekta Unity, Philips Marlin Software versions: R5.3.30, R5.

Recalled Product

Elekta Unity, Philips Marlin Software versions: R5.3.30, R5.3.31, R6.0.5331, R5.7.0, R5.7.1, R6.1.571


Hazard / Issue

In TSM (Treatment Session Manager) Motion Monitoring workflows, under certain conditions there can be a mismatch between the contour data overlay with respect to the motion monitoring images of the monitored structure.

Issued by

FDA

Affected States: IA, NY, PA, TX, WI
Lot/Code Info: Software versions: R5.3.30, R5.3.31, R6.0.5331, R5.7.0, R5.7.1, R6.1.571. UDI Number 05060191071321. Anatomy (Protocol Name): Abdomen (btFFE Cor RealTime, btFFE Cor Sag RealTime, btFFE Cor Sag Tra RealTime, btFFE Fast Sag Cor RealTime, btFFE Sag RealTime); HeadNeck L (bFFE Cor RealTime, bFFE Cor Sag RealTime, bFFE Cor Sag Tra RealTime, bFFE Sag RealTime, bFFE Sag Tra RealTime); HeadNeck M (bFFE Cor RealTime, bFFE Cor Sag RealTime, bFFE Cor Sag Tra RealTime, bFFE Sag RealTime, bFFE Sag Tra RealTime); Pelvis L (bFFE Cor RealTime, bFFE Cor Sag RealTime, bFFE Cor Sag Tra RealTime, bFFE Sag RealTime, bFFE Sag Tra RealTime); Pelvis M (bFFE Cor RealTime, bFFE Cor Sag RealTime, bFFE Cor Sag Tra RealTime, bFFE Sag RealTime, bFFE Sag Tra RealTime); Thorax (btFFE Cor RealTime, btFFE Cor Sag RealTime, btFFE Cor Sag Tra RealTime, btFFE Sag RealTime, btFFE Sag RealTime)