medical Class III Updated 2021-08-25

Cell Marque Corporation recalls Cell Marque Antibodies, PSA (ER-PR8) Mouse Monoclonal Antibo

Recalled Product

Cell Marque Antibodies, PSA (ER-PR8) Mouse Monoclonal Antibody, REF: 324M-15, 324M-16


Hazard / Issue

Mouse Monoclonal Antibody labelled with an incorrect expiry date of 31 Dec 2023. The correct expiry date is 31 Dec 2022. If used beyond expiry, the material performance is likely to decrease and may be observed as weak or no staining, and there is a potential risk that a false negative result could be recorded and may lead to delay of diagnosis due to retest.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Item/Lot: 324M-15/0000116695, 324M-16/0000116696
View official government recall