GE Medical Systems, LLC recalls Allia IGS 3 Pulse angiographic X-ray system
See all recalls from GE Medical Systems, LLC →Recalled Product
Allia IGS 3 Pulse angiographic X-ray system
Hazard / Issue
GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.
Issued by
FDA
Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI: 00195278719577; Serial Numbers: M3-24-053, M3-24-124, M3-25-029, M3-25-051, M3-25-053, M3-25-091, M3-25-098, M3-25-124, M3-26-010, M3-26-030