medical Class II Updated 2021-09-01

Philips North America Llc recalls Ingenia 1.5T-Magnetic Resonance Medical Electrical Systems i

Recalled Product

Ingenia 1.5T-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body or extremities Model Number: 781341


Hazard / Issue

Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: S/N and UDI: 84344 (01)00884838055322(21)84344; 87019 (01)00884838055322(21)87019; 87231 (01)00884838055322(21)87231. OUS: 87171 (01)00884838055322 84943 (01)00884838055322 87142 (01)00884838055322 87249 (01)00884838055322 87252 (01)00884838055322 87195 (01)00884838055322 87129 (01)00884838055322 87136 (01)00884838055322 87175 (01)00884838055322 87196 (01)00884838055322 87197 (01)00884838055322 87235 (01)00884838055322 87187 (01)00884838055322 87212 (01)00884838055322 84572 (01)00884838055322 87116 (01)00884838055322 87152 (01)00884838055322 87161 (01)00884838055322 87237 (01)00884838055322 87244 (01)00884838055322 87090 (01)00884838055322 87093 (01)00884838055322 87259 (01)00884838055322 87209 (01)00884838055322