medical Class II Updated 2023-08-16

Waldemar Link GmbH & Co. KG (Mfg Site) recalls LINK MP Monoblock Hip Stem, Trial Stem Instruments, Size 14-

Recalled Product

LINK MP Monoblock Hip Stem, Trial Stem Instruments, Size 14-Indicated for patients with Mobility-limiting diseases, fractures or defects of the hip joint or proximal femur. Item Number: 136-114/00


Hazard / Issue

Increased force may be required to remove the LINK MP Monoblock trial hip stem (Sizes 14mm - 25mm),after impaction. Potential for prolongation of surgery or the surgical procedure may have to be modified due to this issue

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI: 04026575182084 Lot Numbers: B923005 B932048
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →